1.Turn on the power of the spectrophotometer
and allow it to warm up thoroughly (at least
30 minutes, and in some instruments, it may
require 1 hour for stability). During the
warm-up time, measure the diameter and
length of the sample tube.
2.Prepare a white paper strip approximately
2 cm wide and slightly smaller than the
diameter of the sample tube (with sufficient
thickness to reflect the incident light from
the wavelength selector). Place it inside an
empty sample tube.
3.Insert the sample tube into the sample
chamber without covering the lid, allowing
observation of the monochromatic light
reflected from the white paper strip inside
the sample tube.
4.Use the wavelength adjustment knob to set
the wavelength to approximately 650 nm (or
another wavelength with a distinct color),
and observe the reflected light inside the
sample tube. Adjust the height or rotation
of the sample tube to enhance the intensity
of the reflected light. If the light is
still too weak, use the 100%T adjustment
knob.
5.Slowly change the wavelength from 400 to
700 nm using the wavelength adjustment knob.
Pay attention to the blue, green, and red
lights, and record their wavelength ranges.
|